10
SENSe Subsystem
Syntax
Parameters
Item
character_data
[a] When a channel is set to NORMal or DOUBle, TRIG:COUNt is set automatically to 1.
Example
MRAT DOUBle
Reset Condition
On reset, the speed is set to NORMal.
Query
[SENSe[1]]|SENSe2:MRAT?
The query returns the current speed setting, either NORMal, DOUBle or FAST.
428
:
SENS
1
SENS2
Description/Defaul t
A numeric value for the measurement speed:
–
: 20 readings/second
NORMal
–
: 40 readings/second
DOUBle
–
up to 1000 readings/second
FAST:
The default is
NORMal
N1911A/1912A P-Series Power Meters Programming Guide
MRAT
Space
.
This command sets the Channel A speed to
40 readings/second.
numeric_value
?
Range of Values
[a]
NORMal
[a]
DOUBle
FAST