Keysight N1911A Programming Manual page 400

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe?
The query returns a 1 or 0 into the output buffer indicating whether a table is
enabled or disabled.
– 1 is returned when the table is enabled
– 0 is returned when the table is disabled
Query Example
SENS2:CORR:CSET1:STAT?
Error Messages
– If you attempt to set this command to ON and no table has been selected using
– If the CSET1 command is used when a P-Series, N8480 Series (excluding
400
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect] then error –221,
"Settings conflict" occurs and
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:STATe remains OFF.
Option CFT) or an E-Series power sensor is connected, the error –241
"Hardware missing" occurs.
N1911A/1912A P-Series Power Meters Programming Guide
This command queries whether there is
currently an active sensor calibration table
for Channel B.