CONFigure[1]|2|3|4[:SCALar][:POWer:AC]:RATio
[<expected_value>[,<resolution>[,<source list>]]]
This command sets the measurement function, range and resolution of the
specified window. It sets the measurement function to ratio with relative mode off.
Syntax
CONF
1
2
3
4
Parameters
Refer to
parameters in this command.
Item
expected_value
(for the expected power level)
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command sub-systems. The parameters must
be entered in the specified order. If parameters are omitted, they default from the right. The parameter DEFault is used as a place holder.
Specifying DEF leaves the parameter value unchanged.
128
:POW
:SCAL
expected_value
Space
DEF
"Optional Parameters"
Description/Defaul t
The power meter ignores the numeric value entered in this
parameter. Any value entered is treated like DEF.
A numeric value for the resolution. If unspecified the
current resolution setting is used.
This channel list specifies the channels used to
calculate the ratio. If unspecified and the current
window setup is a ratio measurement then this ratio
setup is used, otherwise it defaults to Channel A/B
(N1912A) or A/A (N1911A).
N1911A/1912A P-Series Power Meters Programming Guide
:AC
:RAT
,
resolution
DEF
on page 112 for additional details on the
,
source list
Range of Values
sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
[c]
(@1),(@2)
[c]
(@2),(@1)
(@1),(@1)
[c]
(@2),(@2)