Example
Error Messages
– INITiate:CONTinuous must be set to OFF on both channels, otherwise error
– If TRIGger:SOURce is set to BUS or HOLD on either channel, error –214, "Trigger
– If the resolution parameter is not the same as the current resolution setting on
N1911A/1912A P-Series Power Meters Programming Guide
READ1:DIFF:REL? DEF,4,
(@2),(@1)
–213, "INIT ignored" occurs.
deadlock" occurs.
the specified window, error –221, "Settings conflict" occurs.
MEASurement Commands
This command queries the upper
window/upper measurement
relative difference measurement of
Channel B -Channel A, using the
current sensor range and a
resolution setting of 4 on both
channels.