14
TRIGger Subsystem
TRIGger[:SEQuence]:HYSTeresis <numeric_value>
This command sets:
– How far a signal must fall below TRIG:LEVel before a rising edge can be
– How far a signal must rise above TRIG:LEVel before a falling edge can be
Syntax
Parameters
Item
numeric_value
664
detected.
detected.
TRIG
:SEQ
Description/Defaul t
How far a signal must fall/rise before a rising or falling edge can
be detected.
– DEF: the default value is 0 dB
Units are resolved to 0.05 dB.
:HYST
Space
N1911A/1912A P-Series Power Meters Programming Guide
numeric_value
DEF
?
Range of Values
0 to 3 dB
DEF