Output:roscillator[:State] <Boolean - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

8
OUTPut Subsystem
OUTPut:ROSCillator[:STATe] <boolean>
This command enables or disables the POWER REF output.
Syntax
OUTP
Example
OUTP:ROSC:STAT 1
Reset Condition
On reset, the POWER REF output is disabled.
Query
OUTPut:ROSCillator[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating whether or
not the POWER REF is enabled.
– 1 is returned when the POWER REF output is enabled
– 0 is returned when the POWER REF output is disabled
Query Example
OUTP:ROSC?
312
:ROSC
:STAT
N1911A/1912A P-Series Power Meters Programming Guide
Space
0|OFF
1|ON
?
This command enables the POWER REF
output.
This command queries the status of the
POWER REF output.

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a