Syntax
SENS
SENS2
Parameters
Item
numeric_value
Example
AVER:COUN 400
Reset Condition
On reset, the filter length is set to 4.
Query
AVERage:COUNt? [MIN|MAX]
The query returns the current setting of the filter length or the values associated
with MIN and MAX. The format of the response is <NR1>.
N1911A/1912A P-Series Power Meters Programming Guide
:
AVER
1
Description/Defaul t
A numeric value defining the filter length.
DEF: the default value is 4
MIN: 1
MAX: 1024
Space
:COUN
?
This command enters a filter length of 400
for Channel A.
SENSe Subsystem
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
Range of Values
1 to 1024
DEF
MIN
MAX
10
371