TRIGger Subsystem

INITiate[1]|2[:IMMediate]

This command sets the power meter in the wait for trigger state. When a trigger is
received, the measurement is taken and the result placed in the power meter
memory. If TRIGger:SOURce is set to IMMediate the measurement begins as
soon as INITiate:IMMediate is executed.
Use FETCh? to transfer a measurement from memory to the output buffer. Refer
to
This command performs the same function as
NOTE
INITiate:[IMMediate]:SEQuence[1]|2.
Syntax
INIT
Example
INIT2:IMM
Error Messages
If the power meter is not in the idle state or INITiate:CONTinuous is ON, error
–213, "INIT ignored" occurs.
646
"FETCh[1]|2|3|4 Queries"
1
2
N1911A/1912A P-Series Power Meters Programming Guide
on page 132 for further details.
:IMM
This command places Channel B in the wait
for trigger state.