Reset Condition
On reset, X Start = 0 s; X Scale = 10 µs.
Query
SENSe[1]|2:TRACe:X:SCALe:PDIV?
The query returns the current scale setting of X-axis in numerical value.
Query Example
SENS:TRAC:X:SCAL:PDIV?
Error Messages
– If no sensor or wrong sensor is connected to the channel, error –241
– If channel is not in trigger acquisition mode, error –221 "Settings conflict"
– If limits of the values keyed in are exceeded, error –222 "Data out of range;
N1911A/1912A P-Series Power Meters Programming Guide
"Hardware missing" occurs.
occurs.
upper (or lower) limit exceeded; no change" occurs.
This command queries the Channel A current
X-axis scale setting in numerical value.
SENSe Subsystem
10
461