Keysight N1911A Programming Manual page 180

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

3
CALCulate Subsystem
Parameters
Item
Description
string
The input measurement type to be fed to the specific input on the CALC block:
– PEAK: peak power
– PTAV: peak to average
– AVER: average
– MIN: minimum power
Values may be followed by ON SWEEP[1]|2|3|4 where the numeric
specifies the gate to be used for the feed. For example: "POW:PEAK ON
SWEEP2".
If ON SWEEP[1]|2|3|4 is not supplied, the gate used is left unchanged.
A feed of "" (empty string) disables the CALC block and switches off that
display line.
Example
CALC3:FEED2 "POW:AVER ON
SWEEP2"
Reset Condition
On reset, data_handle is set to :POW:AVER.
Query
CALCulate[1]|2|3|4:FEED[1]2?
The query returns the current value of the string.
180
This command selects the input for FEED2
of CALC block CALC3 to be average power,
using gate 2. The channel from which the
feed is taken is determined by
CALC:MATH:EXPR..
N1911A/1912A P-Series Power Meters Programming Guide
Range of Values
"POW:PEAK"
"POW:PTAV"
"POW:AVER" "POW:MIN"