Table 1-7
Bit Definitions - Questionable Status Registers
Bit Number
Decimal Weight
0 to 2
-
3
8
4 to 7
-
8
256
9
512
10 to 14
-
15
-
The condition bits are set and cleared under the following conditions:
Table 1-8
Bit change conditions for Questionable Status Register
Bit Number
Meaning
3
POWer Summary
N1911A/1912A P-Series Power Meters Programming Guide
Definition
Not used
POWer Summary
Not used
CALibration Summary
Power On Self Test
Not Used
Not used (always 0)
EVENts Causing Bit Changes
This is a summary bit for the Questionable POWer Register.
– SET:
Error –230, "Data corrupt or stale"
Error –231, "Data questionable;Input Overload"
Error –231, "Data questionable;Input Overload ChA"
Error –231, "Data questionable;Input Overload ChB"
Error –231, "Data questionable;PLEASE ZERO"
Error –231, "Data questionable;PLEASE ZERO ChA"
Error –231, "Data questionable;PLEASE ZERO ChB"
Error –231, "Data questionable;Lower window log error"
Error –231, "Data questionable;Upper window log error"
– CLEARED: When no errors are detected by the power meter during a
measurement covering the causes given for it to set.
Power Meter Remote Operation
[a]
[a]
[a]
[a]
[a]
[a]