PSTatistic[1]|2:CCDF:DATa?

This command is used to return 501 probability values in % at different power
level within certain range, starts from 0 dB till the predefined maximum power
level.
The maximum power level can be set by using this command:
NOTE
PSTatistic[1]|2:CCDF:DATa:MAX <numeric_value>
By default, the maximum value is 50 dB.
The power interval between each reading (probability value) is determined by
the defined maximum power level divided by 500.
This command is only applicable when P-Series sensors are present and the
NOTE
following conditions are met:
– Free run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen
Syntax
Example
PST1:CCDF:DAT?
N1911A/1912A P-Series Power Meters Programming Guide
:CCDF
1
PST
2
PSTatistic Subsystem
:DAT
?
This command returns 501 probability values
in % at different power levels within certain
range (from 0 dB to maximum power level
defined).
9
343