Setting Measurement Limits

You can configure the power meter to detect when a measurement is outside of a
predefined upper and/or lower limit value.
Limits are window or measurement display line based and can be applied to
power, ratio or difference measurements.
Setting Limits
The power meter can be configured to verify the power being measured against
an upper and/or lower limit value. The range of values that can be set for lower
and upper limits is –150.00 dBm to +230.00 dBm. The default upper limit is
+90.00 dBm and the default lower limit is –90.00 dBm.
A typical application for this feature is shown in
Figure 1-4
N1911A/1912A P-Series Power Meters Programming Guide
Swept Source
OUT
Limits Checking Application
Power Meter Remote Operation
Figure
Device
Under Test
IN
OUT
1-4.
Power Meter
CHANNEL A
INPUT
1
63