Keysight N1911A Programming Manual page 761

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

E4410 Series Sensors:
Calibration Factor Block
Layout
These table entries are repeated as shown for each frequency point
Frequency (point 'N')
[a]
Cal factor (low power)
[a]
Cal factor (high power)
Table Size:
Table A-2
Calibration Factor Block Layout: E9300 Series Sensors
E9300 Series Sensors:
Calibration Factor Block
Layout
Header:
Number of tables
Number of frequency
points
Bytes per frequency point 1
Frequency LSB weight
N1911A/1912A P-Series Power Meters Programming Guide
No. Bytes
Contents
4
-
2
-
2
-
-
[a]
See note
No. Bytes
Contents
1
2
2
-
-
2
1000
Calibration Factor Block Layout
Data Format
Data Range
32 bit fixed
0 to Fhpb*
(2^32)
2.14
0.25 to 3
2.14
0.25 to 3
Data Format
Data Range
16 bit integer
-
-
-
Units
Notes
None
Fhbp = Freq Hz per bit
None
Power (in watts) is divided
by this value.
None
Power (in watts) is divided
by this value.
The table size is
dependent on the number
of frequency points.
Units
Notes
None
Number of cal factor
tables. Note that the
power levels and
frequency points are the
same for all tables.
None
None
Number of bytes in cal
factor value at each
frequency and power level.
Hertz
Fhbp (Freq. Hz per bit).
1 KHz per bit for the cal
factor: 1KHz x 2^32 =
4.3E+12 = 4300 GHz
range.
A
761