8
OUTPut Subsystem
Query
OUTPut:TRIGger[:STATe]?
The query command enters a 1 or 0 into the output buffer indicating whether or
not the trigger output signal is enabled or disabled.
– 1 is returned when the trigger output signal is enabled
– 0 is returned when the trigger output signal is disabled
Query Example
OUTP:TRIG:STAT?
Error Messages
– If 8480, N8480, E4410, E9300 or E9320 sensor is connected and the trigger
314
source is not set to external, error –221 "Settings conflict" occurs.
N1911A/1912A P-Series Power Meters Programming Guide
This command queries the status of the
trigger output signal.