[Sense[1]]|Sense2:Correction:cfactor|Gain[1][:Input] [:Magnitude] <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:CORRection:CFACtor|GAIN[1][:INPut]
[:MAGNitude] <numeric_value>
This command is used to enter a gain correction value for the calibration factor.
The power meter corrects every measurement by this factor to compensate for the
gain.
Either CFACtor and GAIN1 can be used in the command—both have an identical
result. Using GAIN1 complies with the SCPI standard, whereas CFACtor does
not—this may make your program easier to understand.
Syntax
SENS
SENS2
N1911A/1912A P-Series Power Meters Programming Guide
:
1
CORR
:GAIN
1
:CFAC
Space
?
SENSe Subsystem
:INP
:MAGN
numeric_value
DEF
MIN
MAX
Space
MIN
MAX
10
393