[Sense[1]]|Sense2:Sweep[1]|2|3|4:Marker[1]|2:Power - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

[SENSe[1]]|SENSe2:SWEep[1]|2|3|4:MARKer[1]|2:POWer?

This query is used to retrieve the power measurement at marker 1 or marker 2
specified for the gate.
The following settings are required to use the command:
– Sensor mode: Peak
– Trigger mode: Single INT/EXT trigger (INIT:CONT OFF) or continuous
Syntax
SENS
SENS2
Example
SENS:SWE2:MARK1:POW?
Error Messages
If the query is used when the following conditions are not met, error –221,
"Settings conflict" occurs:
– Sensor mode: Peak
– Trigger mode: Single INT/EXT trigger (INIT:CONT OFF) or continuous
N1911A/1912A P-Series Power Meters Programming Guide
INT/EXT trigger (INIT:CONT ON)
:
SWE
1
INT/EXT trigger (INIT:CONT ON)
:MARK
1
1
2
2
3
4
This query returns the power measurement
at marker 1 of Gate 2 of Channel A.
SENSe Subsystem
:POW
?
10
439