Calculate[1]|2|3|4:Gain:state <Boolean - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

CALCulate[1]|2|3|4:GAIN:STATe <boolean>
This command is used on the specified window/measurement to enable and
disable the display offset set by the CALCulate[1]|2|3|4:GAIN[:MAGNitude]
command.
Syntax
Example
CALC2:GAIN:STAT 1
Reset Condition
On reset, the gain is disabled.
Query
CALCulate[1]|2|3|4:GAIN:STATe?
The query enters a 1 or 0 into the output buffer indicating the status of the display
offset.
– 1 is returned when the display offset feature is enabled
– 0 is returned when the display offset feature is disabled
N1911A/1912A P-Series Power Meters Programming Guide
CALC
1
2
3
4
:STAT
:GAIN
This command enables the display offset
for the lower window/ upper measurement.
CALCulate Subsystem
Space
0|OFF
1|ON
?
3
185

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a