3
MEASurement Commands

CONFigure [1]|2|3|4 Commands

The CONFigure command is used on the specified window/measurement to set:
– the expected power level being measured.
– the resolution of the window/measurement.
– the channel on which the measurement is to be made.
The expected power level parameter is only required for non-thermistor sensors.
NOTE
This parameter is provided for backward compatibility purpose and has no effect
on thermistor sensors.
The CONFigure command does not make the power measurement after setting
the configuration. Use READ?, or alternatively use INITiate followed by a FETCh?
to make the measurement.
The CONFigure command also applies the following default settings in the
specified window.
Defaul t settings
INITiate:CONTinuous OFF
TRIGger:SOURce IMMediate
TRIGger:DELay:AUTO ON
152
Description
Sets the N432A to make one trigger cycle when INITiate is sent
Sets the N432A to make the measurement immediately once a trigger is received
Enables automatic delay before making the measurement