Initiate:continuous:all <Boolean - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

INITiate:CONTinuous:ALL <boolean>
Sets all trigger sequences to be continuously initiated.
If INITiate:CONTinuous:ALL is set to:
– ON, trigger sequences are set to be continuously initiated
– OFF, trigger sequences are not set to be continuously initiated
Syntax
Example
INIT:CONT:ALL ON
Reset Condition
On reset (*RST), this command is set to OFF.
On preset (SYSTem:PRESet) and instrument power-up, when entering local
mode, if TRIGger[:SEQuence[1]|2]:SOURce is set to INT[[1]|2] or EXT,
INITiate:CONTinuous is not changed. For other trigger sources,
INITiate:CONTinuous is set to ON.
N1911A/1912A P-Series Power Meters Programming Guide
INIT
:CONT
Space
:ALL
This command sets all trigger sequences to
be continuously initiated.
TRIGger Subsystem
0|OFF
1|ON
?
14