Figure 1-7 Generalized Status Register Model - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

1
Power Meter Remote Operation
Bit 0
Bit 1
Bit 2
Bit 3
Figure 1-7
When a status group is implemented in an instrument, it always contains all of the
component registers. However, there is not always a corresponding command to
read or write to every register.
Condition Register
The condition register continuously monitors the hardware and firmware status of
the power meter. There is no latching or buffering for this register, it is updated in
real time. Condition registers are read-only.
Transition Filter
The transition filter specifies which types of bit state changes in the condition
registers and set corresponding bits in the event register. Transition filter bits may
be set for positive transitions (PTR), negative transitions (NTR), or both. Transition
filters are read-write. They are unaffected by *CLS or queries. After
STATus:PRESet the NTR register is set to 0 and all bits of the PTR are set to 1.
Event Register
The event register latches transition events from the condition register as
specified by the transition filter. Bits in the event register are latched and on
setting they remain set until cleared by a query or a *CLS. Also on setting, an
event bit is no longer affected by condition changes. It remains set until the event
register is cleared; either when you read the register or when you send the *CLS
(clear status) command. Event registers are read-only.
74
Condition
Transition
Register
Filter
0
1
2
Generalized Status Register Model
N1911A/1912A P-Series Power Meters Programming Guide
Event
Enable
Register
Register
Summary