Query Example
SENS2:AVER?
Error Messages
– If [SENSe[1]]|SENSe2:AVERage:STATe is set to ON while
– If [SENSe[1]]|SENSe2:AVERage:STATe is set to ON when a N1920 or E9320
N1911A/1912A P-Series Power Meters Programming Guide
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, "Settings Conflict"
occurs.
power sensor is connected in AVERage measurement mode and is in the
wait-for-trigger state for external trigger buffering, the error –221, "Settings
Conflict" occurs.
This command queries whether averaging is
on or off for Channel B.
SENSe Subsystem
10
379