Table 7-3
Frequency and Calibration/Offset Factor List
Table
Power Sensor
0
None
1
8481A
2
8482A, 8482B, 8482H
3
8483A
4
8481D
5
8485A
6
R8486A
7
Q8486A
8
R8486D
9
8487A
[a] Default is a sensor calibration table in which the reference calibration factor and calibration factors are 100%. This sensor calibration table
can be used during the performance testing of the power meter.
For sensosr calibration tables (8480 Series sensors and N8480 Series sensors with
Option CFT only), the number of frequency points must be one less than the
number of calibration factor points. This is verified when the sensor calibration
table is selected using SENSe:CORRection:CSET:SELect <string>.
Ensure that the frequency points you use cover the frequency range of the signals
that you want to measure. If you measure a signal with a frequency outside the
frequency range defined in the table, then the power meter uses the highest or
lowest point in the table to calculate the calibration factor/offset.
Depending on available memory, the power meter is capable of storing 20 sensor
calibration tables and 10 frequency dependent offset tables, each containing
80 points.
N1911A/1912A P-Series Power Meters Programming Guide
MEMory Subsystem
Table Name
[a]
DEFAULT
8481A
8482A
8483A
8481D
8485A
R8486A
Q8486A
R8486D
8487A
7
291