[Sense[1]]|Sense2:Power:ac:range <Numeric_Value - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

10
SENSe Subsystem
[SENSe[1]]|SENSe2:POWer:AC:RANGe <numeric_value>
This command is only valid when used with an E-Series power sensor. Its purpose
is to select one of two power ranges.
– If 0 is selected, the power sensor's lower range is selected
– If 1 is selected, the power sensor's upper range is selected
Setting a range with this command automatically switches
[SENSe[1]]|SENSe2:POWer:AC:RANGe:AUTO to OFF.
Syntax
SENS
SENS2
Example
POW:AC:RANG 0
Reset Condition
On reset, the upper range is selected.
Query
[SENSe[1]]|SENSe2:POWer:AC:RANGe?
The query enters a 1 or 0 into the output buffer indicating the status of the power
sensor's range.
– 1 is returned when the upper range is selected
– 0 is returned when the lower range is selected
430
:
POW
1
N1911A/1912A P-Series Power Meters Programming Guide