12
SYSTem Subsystem
Command
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate Setup
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Trigger setup
TRIG[:SEQ[1]|2]:SOUR
INIT:CONT
TRIG[:SEQ]:LEV:AUTO
TRIG[:SEQ]:LEV
TRIG[:SEQ]:SLOP
TRIG[:SEQ]:DEL
TRIG[:SEQ]:HOLD
TRIG[:SEQ]:HYST
Range
[SENS[1]]|SENS2:POW:AC:RANG:AUTO
[SENS[1]]|SENS2:POW:AC:RANG
Averaging
[SENSe[1]]|SENSe2:AVER[:STATe]
[SENSe[1]]|SENSe2:AVER:COUN
Step detection
538
Setting
NORM
E9321A/25A: HIGH
E9322A/26A: MED
E9323A/27A: LOW
N1921/2A: LOW
Gate 1: 20 µs
Gates 2 - 4: 0
Gate 1: 520 µs
Gates 2 - 4: 0
INT1
ON
OFF
–15 dBm
POS
0 s
4275 µs
3 dB
OFF
UPPER
ON
64
N1911A/1912A P-Series Power Meters Programming Guide
Comments
Measurement mode
Sensor video bandwidth
Delay between trigger point and time gated
period.
Length of time gated period for time gated
measurements.
Trigger source set up and acquisition mode
continuous triggering
Disable automatic setting of the trigger
level
Power level
Trigger event recognized on the rising edge
of a signal
Delay between recognition of trigger event
and start of a measurement
Trigger holdoff
Hysteresis
Auto range off
Range set to upper
Averaging On
Averaging set to 64