TRIGger Subsystem
TRIGger[:SEQuence[1]|2]:COUNt <numeric_value>
This command controls the path of the trigger subsystem in the upward traverse
of the wait for trigger state. COUNt loops through the event detection/
measurement cycle are performed. That is, COUNt measurements are performed in
response to COUNt trigger events.
COUNt can be set to a value >1 only when:
– [SENSe[1]]|SENSe2:MRATe <character_data> is set to FAST
– TRIGger[1]|2:SOURce set to BUS, IMMediate or HOLD.
When COUNt is set to a value >1,
– CALibration[1]|2:ZERO:AUTO will switch to OFF automatically. It will
– Setting a channel from FAST mode to NORMal mode or DOUBle mode will also
Syntax
TRIG
672
restored to its default setting when the COUNt is set to 1.
restore both the CALibration[1]|2:ZERO:AUTO and COUNt to its default
setting automatically.
:SEQ
N1911A/1912A P-Series Power Meters Programming Guide
1
:COUN
2
numeric_value
Space
DEF
?