Keysight N1911A Programming Manual page 333

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Query
PSTatistic:CCDF:REFerence[:STATe]?
The query enters a 1 or 0 into the output buffer indicating the status of the
reference trace stored.
– 1 is returned when the previously stored reference trace is turned on
– 0 is returned when the previously stored reference trace is turned off
Query Example
PST:CCDF:REF?
Error Messages
– If command is executed in other window besides expanded CCDF window,
– If there was no previously saved trace, error –221 "Settings conflict: No
N1911A/1912A P-Series Power Meters Programming Guide
error –221 "Settings conflict: Requires CCDF expanded window" occurs.
reference trace saved" occurs. Please check the status of saved reference trace
using command PST:CCDF:STOR:REF?.
PSTatistic Subsystem
This command queries whether the reference
trace is turned on or off.
9
333

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents