Example
CORR:CSET1 'PW1'
Reset Condition
On reset the selected table is not affected.
Query
[SENSe[1]]|SENSe2:CORRection:CSET[1]|CSET2:[SELect]?
The name of the selected table is returned as a quoted string. If no table is
selected an empty string is returned.
Query Example
CORR:CSET1?
Error Messages
– If <string> is not valid, error –224, "Illegal parameter value" occurs.
– If a table called <string> does not exist, error –256, "File name not found"
– When a sensor calibration table is selected, the power meter verifies that the
– If the CSET1 command is used when a P-Series or an E-Series power sensor is
398
occurs.
number of calibration points defined is one more than the number of frequency
points defined. When a frequency dependent offset table is selected, the
power meter verifies that the number of offset points defined is equal to the
number of frequency points defined. If this is not the case, error –226, "Lists
not the same length" occurs.
connected the error –241, "Hardware missing" occurs.
N1911A/1912A P-Series Power Meters Programming Guide
This command enters the name of the sensor
calibration table which is to be used on
Channel A.
This command queries the sensor calibration
table currently used for Channel A.