SENSe[1]|SENSe2:TRACe:OFFSet:TIME <numeric_value>
This command sets the delay between the delayed trigger point and the start of
the trace for P-Series or E9320 sensors are set to NORMal mode.
To set an E9320 sensor to NORMal mode, refer to the command
"[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>"
Syntax
SENS
SENS2
Parameters
Item
numeric_value
Example
SENS:TRAC:OFFS:TIME 0.05
450
:
TRAC
1
Description/Defaul t
The length of the delay in seconds.
: the default value is 0 seconds
DEF
Units are resolved to 1 ns.
N1911A/1912A P-Series Power Meters Programming Guide
:OFFS
Space
:TIME
This command sets the delay to 0.05
seconds.
on page 414.
numeric_value
DEF
?
Range of Values
–1 to 1 second
DEF