Parameters
Item
numeric_value
Example
SENS2:TRAC:LIM:LOW 10
Reset Condition
On reset, the value is set to –50 dBm.
Query
SENSe[1]|2:TRACe:LIMit:LOWer [MIN|MAX]
The query returns the current setting of the trace lower scale limit or the value
associated with MIN or MAX. The format of the response is <NR3>.
Query Example
SENSe:TRAC:LIM:LOW?
N1911A/1912A P-Series Power Meters Programming Guide
Description/Defaul t
A numeric value for the trace lower scale limit.
: the default is 20 dBm
DEF
: –150 dBm
MIN
: 230 dBm
MAX
Range of Values
–150 to 230 dBm
DEF
MIN
MAX
This command sets the trace lower scale
limit to 10 dBm for Channel B.
This command queries the trace lower scale
limit of Channel A.