9
PSTatistic Subsystem
PSTatistic:CCDF:MARKer[1]|2:Y <numeric_value>
This command is used to set the current marker Y-axis position at the selected
trace.
This command is only applicable when P-Series sensors are present and the
NOTE
following conditions are met:
– Free-run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen
Syntax
PST
Parameter
Item
numeric_value
[a] The marker will be placed at the point nearest to the specified Y-axis position if that particular Y value is not available.
328
:CCDF
:MARK
Description/Defaul t
The current marker Y-axis position
– Minimum Value: 0 %
– Maximum Value: 100 %
N1911A/1912A P-Series Power Meters Programming Guide
1
Space
:Y
2
[a]
.
numeric_value
Range of Values
0 to 100