N8480 Series Sensors:
Calibration Factor Block
Layout
[a]
Cal factor (high power)
Table Size:
[a] Corrections are applied in power for E4410 Series, E9300 Series and N8480 Series sensors (excluding Option CFT).
[b] Corrections are applied in voltage versus ADC reading for E9320 Series sensors. This format also requires only one correction factor across all
power levels.
[c] The block layout shown for E9320 Series sensors exists in two separate EEPROM locations. One location contains the calibration factor data
for the average path and the other contains the calibration factor data for the peak path. These EEPROM blocks are accessed using the
SERV:SENS:CALFactor and SERV:SENS:PCALfactor commands respectively.
N1911A/1912A P-Series Power Meters Programming Guide
No. Bytes
Contents
2
-
-
[a]
See note
Calibration Factor Block Layout
Data Format
Data Range
2.14
0.25 to 3
Units
Notes
None
Power (in watts) is divided
by this value.
The table size is
dependent on the number
of frequency points.
A
765