Device Status Register Sets

The status registers contain information which give device status information. The
contents of the individual registers of these register sets may be accessed by
appending the commands listed in
The following command descriptions detail the SCPI register you require to
control but do not detail the register set commands.
The one device status register set is:
STATus:DEVice:
The following bits in these registers are used by the power meter:
Bit Number
Decimal Weight
0
-
1
2
2
4
3
8
4
16
7-15
-
14
16384
15
-
The Channel A and B sensor connected bits (bits 1 and 2), when queried with the
STATus:DEVice:CONDition? query are set to:
– 1, when a power sensor is connected
– 0, when no power sensor is connected
The Channel A and B sensor connected bits (bits 1 and 2), when queried with the
STATus:DEVice:EVENt? query indicate whether a power sensor has been
connected or disconnected depending on the state of the corresponding bits of
STATus:DEVice:NTRansition and STATus:DEVice:PTRansition. If the
corresponding bit in:
N1911A/1912A P-Series Power Meters Programming Guide
Status Register Set
Definition
Not used
Channel A sensor connected
Channel B sensor connected (N1912A only)
Channel A sensor error
Channel B sensor error (N1912A only)
Not used
Front Panel key press
Bit 15 always 0
STATus Subsystem
Commands.