14
TRIGger Subsystem

TRIGger[:SEQuence[1]|2]:IMMediate

This command provides a one time over-ride of the normal process of the
downward path through the wait for trigger state. It causes the immediate exit of
the event detection layer if the trigger system is in this layer when the command is
received. In other words, the instrument stops waiting for a trigger and takes a
measurement ignoring any delay set by TRIG:DELay.
Syntax
Example
TRIG:SEQ:IMM
676
TRIG
:SEQ
1
:IMM
2
This command initiates a measurement on
Channel A.
N1911A/1912A P-Series Power Meters Programming Guide