Keysight N1911A Programming Manual page 375

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

The query enters a 1 or 0 into the output buffer indicating whether automatic filter
length is enabled or disabled.
– 1 is returned when automatic filter length is enabled
– 0 is returned when automatic filter length is disabled
Query Example
AVER:COUN:AUTO?
Error Messages
If [SENSe[1]]|SENSe2:AVERage:COUNt:AUTO is set to ON while
[SENSe[1]]|SENSe2:SPEed is set to 200, the error –221, "Settings Conflict"
occurs. However, automatic averaging is enabled but the
[SENSe[1]]|SENSe2:AVERage:STATe command is not automatically set ON.
N1911A/1912A P-Series Power Meters Programming Guide
This command queries whether automatic filter
length selection is on or off for Channel A.
SENSe Subsystem
10
375