Pstatistic:ccdf:gaussian:marker[1]|2[:Set] - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

PSTatistic:CCDF:GAUSsian:MARKer[1]|2[:SET]

This command is used to set the markers on Gaussian trace. The markers will be
set only if the trace is present and visible. According to the selections made, the
markers will become active on the screen.
This command is only applicable when P-Series sensors are present and the
NOTE
following conditions are met:
– Free-run acquisition mode is selected
– NORMal or DOUBle measurement speed setting is chosen
– Expanded window is enabled (not full CCDF screen window)
Syntax
Example
PST:CCDF:GAUS:MARK
Reset Condition
On reset, the marker is set on the next visible trace according to the sequence
Channel A, Channel B, Reference or Gaussian.
N1911A/1912A P-Series Power Meters Programming Guide
PST
:CCDF
:MARK
:GAUS
This command sets marker 1 on the
Gaussian trace.
PSTatistic Subsystem
1
:SET
2
9
321

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents