SENSe[1]|SENSe2:TRACe:TIME <numeric_value>
This command sets the duration of the trace for a P-Series sensor and a E9320
sensors set to NORMal mode. To set the E9320 sensor to NORMal mode, refer to the
command
page 414.
Syntax
Parameters
Item
numeric_value
Example
SENS2:TRAC:TIME 0.5
452
"[SENSe[1]]|SENSe2:DETector:FUNCtion <character_data>"
:
1
SENS
SENS2
Description/Defaul t
The duration of the trace in seconds.
: the default value is 100 µs.
DEF
Units are resolved to 1 ns.
N1911A/1912A P-Series Power Meters Programming Guide
Space
TRAC
:TIME
This command sets the duration of the trace
to 0.5 seconds for Channel B.
on
numeric_value
DEF
?
Range of Values
20 ns to 1 s
DEF