Keysight N1911A Programming Manual page 202

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

3
CALCulate Subsystem
[a] Quotes are mandatory. Either single or double quotes may be used.
[b] N1912A only.
[c] The mathematical operation will be performed in linear scale.
Example
CALC2:MATH "(SENS2/SENS1)"
Reset Condition
On reset, the Keysight N1911A upper and lower window measurements are set to
Channel A ("(SENS1)"). On the N1912A the upper window measurements are set to
Channel A ("(SENS1)") and the lower window measurements to Channel B ("(SENS2)").
202
This command sets the lower window/
upper measurement to make a Channel
B/A ratio measurement.
N1911A/1912A P-Series Power Meters Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents