Status:operation:sense[:Summary] - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

11
STATus Subsystem

STATus:OPERation:SENSe[:SUMMary]

The operation status sense summary register set contains information on the
status of the power sensors.
The following bits in these registers are used by the power meter:
Bit Number
Decimal Weight
0
-
1
2
2
4
3-15
-
These bits are set when the power meter is reading data from the E-Series power
sensor or N8480 Series power sensor EEPROM.
These bits are cleared when the power meter is not reading data from the E-Series
power sensor or N8480 Series power sensor EEPROM.
Syntax
480
Definition
Not used
Channel A SENSe Status
Channel B SENSe Status (N1912A only)
Not used
:OPER
STAT
:SENS
:SUMM
N1911A/1912A P-Series Power Meters Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents