5
DISPlay Subsystem
DISPlay[:WINDow[1]|2]:ANALog:LOWer <numeric_value>
This command sets the analog meter lower scale limit.
This command has the same purpose as DISPlay[:WINDow[1]|2]:METer:LOWer
NOTE
<numeric_value>.
The units used are dependent on the current setting of UNIT:POWer and
CALCulate:RELative:STATe as shown in
Table 5-1
Measurement Units
Measurement Mode
Single Channel
Ratio
Difference
236
Measurement Type
Avg, Pk
Pk–Avg
Avg, Pk, Pk–Avg
Avg, Pk
Pk–Avg
N1911A/1912A P-Series Power Meters Programming Guide
Table
CALC:REL:STAT OFF
Linear
Log
Watt
dBm
%
dB
%
dB
Watt
dBm
%
dB
5-1.
CALC:REL:STAT ON
Linear
%
%
%
%
%
Log
dB
dB
dB
dB
dB