CALibration Subsystem

CALibration[1]|2[:ALL]

This command is identical to CALibration[1]|2[:ALL]?, however, unlike the
NOTE
query it does not provide a response to indicate whether the calibration has
been successful or not.
This command causes the power meter to perform a calibration sequence on the
specified channel. The command assumes that the power sensor is connected to
the POWER REF output. The calibration sequence consists of:
1 Zeroing the power meter (CALibration:ZERO:AUTO ONCE), and
2 Calibrating the power meter (CALibration:AUTO ONCE).
For 8480 Series power sensors and N8480 Series power sensors with Option CFT,
the reference calibration factor used during this calibration can be derived from
either an active sensor calibration table or the value entered using
CALibration:RCFactor. The actual value used is the one which was most recently
set. That is, a value entered using CALibration:RCFactor is overridden if a sensor
calibration table is subsequently selected and enabled. Conversely,
CALibration:RCFactor overrides any reference calibration factor previously set
from a sensor calibration table. To determine the currently set reference
calibration factor use CALibration:RCFactor?.
E-Series power sensors and N8480 Series power sensors (excluding Option CFT)
have their sensor calibration tables stored in EEPROM which means that the
reference calibration factor is automatically downloaded by the power meter.
214