Trace[1]|2[:Data]? <Character_Data - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

13
TRACe Subsystem
TRACe[1]|2[:DATA]? <character_data>
This query returns trace data from the specified channel. The trace resolution is
determined by <character_data>.
Data is returned in IEEE 488.2 arbitrary block program data format as follows:
#xyyy..yddd....................ddd<LF>
Example: if there are 12435 data bytes, y = 12435 and x = 5
Each point in the trace is represented as an IEEE 754 32 bit floating point number,
made up of four bytes in the data block. The MS byte is transmitted first. Each
complete block is terminated by a line feed.
TRACe data formatting is not affected by FORMat subsystem formatting.
NOTE
Syntax
614
The number of data bytes (d)
contained in the block.
The number of y digits
Signifies the start of the block
TRAC
1
2
N1911A/1912A P-Series Power Meters Programming Guide
?
:DATA
Line feed character
signifies the end of the block
Data bytes
Space
character_data