11
STATus Subsystem
STATus:OPERation
The operation status register set contains conditions which are a part of the
operation of the power meter as a whole.
The following bits in these registers are used by the power meter:
Bit Number
0
1 - 3
4
5
6 - 9
10
11
12
13 to 15
Syntax
STAT
476
Decimal Weight
1
-
16
32
-
1024
2048
4096
-
:OPER
N1911A/1912A P-Series Power Meters Programming Guide
Definition
CALibrating Summary
Not used
MEASuring Summary
Waiting for TRIGger Summary
Not used
SENSe Summary
Lower Limit Fail Summary
Upper Limit Fail Summary
Not used (bit 15 always 0)