Keysight N1911A Programming Manual page 661

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Example
TRIG:SEQ:DEL 0.001
Reset Condition
On reset, the trigger delay is set to 0 seconds.
Query
TRIGger[:SEQuence]:DELay?
The query returns the current setting of the trigger delay.
Query Example
TRIG:SEQ:DEL?
Reset Condition
On reset, trigger delay is set to 0 seconds.
N1911A/1912A P-Series Power Meters Programming Guide
This command sets a delay of 1 ms for
Channel A.
This command queries the trigger delay of
Channel A.
TRIGger Subsystem
14
661

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1912a

Table of Contents