Editing Frequency Dependent Offset Tables - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

1
Power Meter Remote Operation

Editing Frequency Dependent Offset Tables

It is not possible to create any additional frequency dependent offset tables.
However, the 10 existing tables can be edited using the MEMory subsystem.
To do this:
1 Select one of the existing tables using:
2 Enter the frequency data using:
3 Enter the offset factors as shown in the table below using:
4 If required, rename the frequency dependent offset table using:
52
MEMory:TABle:SELect <string>
For information on naming frequency dependent offset tables see
Frequency Dependent Offset Tables"
current names which you can select refer to
Offset Table Names"
MEMory:TABle:FREQuency <numeric_value> {,<numeric_value>}
MEMory:TABle:GAIN <numeric_value> {,<numeric_value>}
Frequency
Frequency 1
Frequency 2
"
Frequency n
MEMory:TABLe:MOVE <string>,<string>. The first <string> parameter
identifies the existing table name, and the second identifies the new table
name.
on page 54. For information on the
on page 53.
N1911A/1912A P-Series Power Meters Programming Guide
"Listing the Frequency Dependent
Offset
Offset 1
Offset 2
"
Offset n