Table 1-10 Bit Change Conditions For Operation Status - Keysight N1911A Programming Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Bit Number
Decimal Weight
10
1024
11
2048
12
4096
13 to 14
-
15
-
The condition bits are set and cleared under the following conditions:
Table 1-10
Bit change conditions for Operation Status
Bit Number Meaning
0
CALibrating
4
MEASuring
5
Waiting for
TRIGger
10
SENSe
11
Lower Limit Fail
12
Upper Limit Fail
N1911A/1912A P-Series Power Meters Programming Guide
Definition
SENSe Summary
Lower Limit Fail Summary
Upper Limit Fail Summary
Not used
Not used (always 0)
EVENts Causing Bit Changes
This is a summary bit for the Operation CALibrating Register.
– SET: At beginning of zeroing (CALibration:ZERO:AUTO ONCE) and at the beginning
of calibration (CALibration:AUTO ONCE). Also for the compound command/query
CALibration[:ALL]?, this bit is set when sensor zeroing begins.
– CLEARED: At the end of zeroing or calibration.
This is a summary bit for the Operation MEASuring Register.
– SET: When the power meter is taking a measurement.
– CLEARED: When the measurement is finished.
This is a summary bit for the Operation TRIGger Register.
– SET: When the power meter enters the "wait for trigger" state.
– CLEARED: When the power meter enters the "idle" state.
This is a summary bit for the Operation SENSe Register.
– SET: When the power meter is reading data from the power sensor's EEPROM.
– CLEARED: When the power meter is not reading data from the power sensor's EEPROM.
This is a summary bit for the Lower Limit Fail Register.
– SET: If a measurement is made and either a channel or window lower limit test fails.
– CLEARED: If a measurement is made and the lower limit test is not enabled or the test is
enabled and passes.
This is a summary bit for the Upper Limit Fail Register.
– SET: If a measurement is made and either a channel or window upper limit test fails.
– CLEARED: If a measurement is made and the upper limit test is not enabled or the test is
enabled and passes.