SERVice:SENSor[1]|2:CALFactor <cal_factor_data>
This command writes calibration factor data to, or reads calibration factor data
from, the currently connected sensor. The whole calibration factor block must be
written at once as a checksum is generated. The new block must not be larger
than the existing block.
This command applies to the following sensors:
– E4410 Series
– N8480 Series (excluding Option CFT)
– E9300 Series
– E9320 Series, average path data
For E9320 Series sensors, peak path, refer to
<cal_factor_data>"
Syntax
SERV
Parameters
Item
cal_factor_data
N1911A/1912A P-Series Power Meters Programming Guide
on page 713.
:SENS
1
2
Description/Defaul t
A binary data block. Refer to
759 for further information.
"SERVice:SENSor[1]|2:PCALfactor
Space
:CALF
Appendix A, "Calibration Factor Block Layout,"
SERVice Subsystem
cal_factor_data
?
on page
15
705