Keyword
:TRANsition[1]|...|10
:NEGative
:POSitive
:REFerence?
:STATe
:UNIT
When making trace measurements, use the following command sequence to
NOTE
synchronize the returned trace data with the measurement:
Command
TRIG:SOUR INT
or
TRIG:SOUR EXT
INIT:CONT OFF
[a]
TRAC:STAT ON
AVER:STAT OFF
or
TRIG:DEL:AUTO OFF
INIT
FETCH?
TRACE:DATA? MRES1
[a] A trace display format must be set when this command is used.
N1911A/1912A P-Series Power Meters Programming Guide
Parameter Form
:DURation?
:OCCurrence?
:DURation?
:OCCurrence?
<numeric_value>
<boolean>
<character_data>
Comment
Change trigger source to internal or external
Trace data can only be retrieved with
INIT:CONT OFF
Enables trace capture
No settling time delays for digital filter to fill
Initiates a new measurement
Fetch the result (waits for the measurement to complete)
Retrieves the trace data once the measurement has completed
TRACe Subsystem
Notes
Page
[query only]
page 628
[query only]
page 629
[query only]
page 630
[query only]
page 631
[query only]
page 632
page 634
page 636
13
613