MEMory Subsystem
MEMory:TABLe:FREQuency <numeric_value>{,<numeric_value>}
This command is used to enter frequency data into the current selected table. Any
previous frequency list is cleared before the new frequency list is stored. The
frequencies must be entered in ascending order. Entries in the frequency lists
correspond as shown in
For sensor calibration tables only, the first calibration factor entered using the
NOTE
MEMory:TABLe:GAIN command is used as the reference calibration factor.
Predefined sensor calibration table is not applicable for N8480 Series power
NOTE
sensors with Option CFT. Therefore you are required to create a new sensor
calibration table for the sensors when a sensor calibration table is needed.
290
Table 7-3
with entries in the calibration/offset factor lists.
N1911A/1912A P-Series Power Meters Programming Guide