PSTatistic[1]|2:CCDF:DATa:MAX <numeric_value>
This command is used to set the maximum value of X-axis CCDF trace.
Syntax
Parameters
Item
numeric_value
Example
PST1:CCDF:DAT:MAX 10
Reset Condition
On reset, the maximum value for CCDF trace X-axis is set to 50 dB.
Query
PSTatistic[1]|2:CCDF:DATa:MAX?
The query returns the X-axis CCDF trace maximum value.
N1911A/1912A P-Series Power Meters Programming Guide
1
:CCDF
PST
2
Description/Defaul t
X-axis CCDF trace maximum value in dB.
– Minimum value: 5.00 dB
– Maximum value: 50.00 dB
PSTatistic Subsystem
:DAT
:MAX
Range of Values
5.00 to 50.00
This command sets the maximum value of
X-axis CCDF trace to 10 dB.
numeric_value
?